Molecular imprinting coupled with electrochemical analysis for plasma samples classification in acute myocardial infarction diagnostic.

Shumyantseva, Victoria V; Bulko, Tatiana V; Sigolaeva, Larisa V; Kuzikov, Alexey V; Pogodin, Pavel V; Archakov, Alexander I.
Biosens Bioelectron; 99: 216-222, 2018 Jan 15.
Article in En | MEDLINE | ID: mdl-28763782

Abstract

Electroanalysis of myoglobin (Mb) in 10 plasma samples of healthy donors (HDs) and 14 plasma samples of patients with acute myocardial infarction (AMI) was carried out with screen-printed electrodes modified first with multi-walled carbon nanotubes (MWCNT) and then with a molecularly imprinted polymer film (MIP), viz., myoglobin-imprinted electropolymerized poly(o-phenylenediamine). The differential pulse voltammetry (DPV) parameters, such as a maximum amplitude of reduction peak current (A, nA), a reduction peak area (S, nA × V), and a peak potential (P, V), were measured for the MWCNT/MIP-sensors after their incubation with non-diluted plasma. The relevance of the multi-parameter electrochemical data for accurate discrimination between HDs and patients with AMI was assessed on the basis of electrochemical threshold values (this requires the reference standard method (RAMP® immunoassay)) or alternatively on the basis of the computational cluster assay (this does not require any reference standard method). The multi-parameter electrochemical analysis of biosamples combined with computational cluster assay was found to provide better accuracy in classification of plasma samples to the groups of HDs or AMI patients.